Hioki SS7012 DC Signal Source
Hioki SS7012 Feature :-
Generate and Measure Signals Simultaneously |
|
· Improve stability and reduce calibration costs compared with the previous HIOKI model · For Instrumentation systems (4 - 20 mA) and loop testing · Check temperature control equipment and electric distribution · 8 types of thermocouples to test thermoelectric power generation · Ideal for electrical device testing and routine maintenance of production equipment such as calibrators · Use the max. 25 mA DC sink as an electric load |
|
Use of the AC Adapter and /or rechargeable batteries and dedicated charger is recommended |
Hioki SS7012 Specification
Basic specifications (Accuracy guaranteed for 1 year)
Circuit method | Bipolar sink and source | Constant Voltage | 2.5 V: 0 to ±2.5000 V (±0.03 % of setting ±300 μV, 100 μV resolution) 25 V: 0 to ±25.000 V (±0.03 % of setting ±3 mV, 1 mV resolution) | Constant Current | 25 mA: 0 to ±25.000 mA (±0.03 % of setting ±3 μA, 1 μA resolution) | Thermoelectric power generation | K: at TC: 0 °C, -174.0 to 1372.0 °C (±0.05 % of setting ±0.5 °C, 0.1 °C resolution), Other types: E, J, T, R, S, B, N selectable | Thermoelectric power generation | K: at TC: RJ, -174.0 to 1372.0 °C (±0.05 % of setting ±1.0 °C, 0.1 °C resolution), Other types: E, J, T, R, S, B, N selectable | Standard resistance (Rs) | 100 Ω (±0.2 Ω) | Automatic generation | Number of memory steps: 20 Interval time: 1 to 99 sec (at CV, CC, TC mode) |
|
|
[Measurement functions] |
Voltage | 2.5 V: 0 to ±2.8000 V (±0.03 % rdg. ±300 μV, 100 μV resolution, 1 MΩ input resistance) 25 V: 0 to ±28.000 V (±0.03 % rdg. ±3 mV, 1 mV resolution, 1 MΩ input resistance) | Current | 25 mA: 0 to ±28.000 mA (±0.03 % rdg. ±3 μA, 1 μA resolution, 25 Ω input resistance) | Temperature | -25.0 to ±80.0 °C (±0.5 °C at 23 ±5 °C, 0.1 °C resolution, use with the RJ sensor 9184) | Sampling rate | Approx. 1.67 times/sec |
|
|
Additional functions | Zero adjustment, Overflow display, USB communication, Monitor | Power supply | AC ADAPTER 9445-02/03 (100 to 240 V AC 50/60 Hz, 9 VA), Ni-MH battery HR6 × 4, 6 VA, (fully charged 2500 mAh Ni-MH batteries: 170 minutes continuous use), or LR6 (AA) alkaline battery × 4, 6 VA | Dimensions, mass | 104 mm (4.09 in) W × 180 mm (7.09 in) H × 58 mm (2.28 in) D, 660 g (23.3 oz) (including LR6 × 4 batteries) | Supplied accessories | INPUT CORD 9168 × 1, TEST LEAD L9170-10 × 1, Fuse × 1, LR6 (AA) alkaline battery × 4, Operation manual × 1 |
|